Nitric oxide (NO)

For our measurement sensor for determining the concentration of Nitric oxide, we offer the following calibration gases.

UsageExplanation
Zero gas This gas is needed for the calibration of the zero point.
Standard test gas The standard test gas is required for calibration / adjustment of the sensor with a two-point calibration. The concentration value corresponds to 70% to 100% of the full scale value of the probe.
Alarm test gas The alarm test gas is required to verify the alarm levels.

 

Select the measuring range of your probe:

 

0 - 100 ppm

Measuring range
of the probe
GasUsageBottle filling volume (Liter)
34 l58 l110 l
0 - 100 ppm Synthetic air Zero gas Z0A0-L1373 Z0A0-M1373 Z0A0-N1373
100 ppm NO in Nitrogen Standard test gas Z0A0-R1613 Z0A0-S1613 Z0A0-T1613
50 ppm NO in Nitrogen Alarm testgas Z0A0-R1610 Z0A0-S1610 Z0A0-T1610
25 ppm NO in Nitrogen Alarm test gas Z0A0-R1606 Z0A0-S1606 Z0A0-T1606

0 - 250 ppm

Measuring range
of the probe
GasUsageBottle filling volume (Liter)
34 l58 l110 l
0 - 250 ppm Synthetic air Zero gas Z0A0-L1373 Z0A0-M1373 Z0A0-N1373
200 ppm NO in Nitrogen Standard test gas Z0A0-R1618 Z0A0-S1618 Z0A0-T1618
100 ppm NO in Nitrogen Alarm test gas Z0A0-R1612 Z0A0-S1612 Z0A0-T1612
50 ppm NO in Nitrogen Alarm test gas Z0A0-R1610 Z0A0-S1610 Z0A0-T1610
25 ppm NO in Nitrogen Alarm test gas Z0A0-R1606 Z0A0-S1606 Z0A0-T1606