n-Butane (C4H10), Methylethylmethane, R 600
For our measurement sensor for determining the concentration of n-Butane, we offer the following calibration gases.
| Usage | Explanation |
|---|---|
| Zero gas | This gas is needed for the calibration of the zero point. |
| Standard test gas | The standard test gas is required for calibration / adjustment of the sensor with a two-point calibration. The concentration value corresponds to about two-thirds of the full scale value of the probe up to 50% LEL. |
| Alarm test gas | The alarm test gas is required to verify the alarm levels. |
Select the measuring range of your probe:
0 - 50% LEL
| Measuring range of the probe | Gas | Usage | Bottle filling volume (Liter) | ||
|---|---|---|---|---|---|
| 34 l | 58 l | 110 l | |||
| 0 - 50% LEL | Synthetic air | Zero gas | Z0A0-L1373 | Z0A0-M1373 | Z0A0-N1373 |
| 0.7 vol% Butane in air | Standard test gas | Z0A0-L0350 | Z0A0-M0350 | Z0A0-N0350 | |
| 0.4 vol% Butane in air | Alarm test gas | Z0A0-L0342 | Z0A0-M0342 | Z0A0-N0342 | |
0 - 100% LEL
| Measuring range of the probe | Gas | Usage | Bottle filling volume (Liter) | ||
|---|---|---|---|---|---|
| 34 l | 58 l | 110 l | |||
| 0 - 100% LEL | Synthetic air | Zero gas | Z0A0-L1373 | Z0A0-M1373 | Z0A0-N1373 |
| 0.7 vol% Butane in air | Standard test gas | Z0A0-L0350 | Z0A0-M0350 | Z0A0-N0350 | |
| 0.4 vol% Butane in air | Alarm test gas | Z0A0-L0342 | Z0A0-M0350 | Z0A0-N0342 | |