i-Butane (C4H10), Isobutane, 2-Methyl propane, Trimethyl methane, 1.1-Dimethyl ethane, R 600a
For our measurement sensor for determining the concentration of i-Butane, we offer the following calibration gases.
| Usage | Explanation |
|---|---|
| Zero gas | This gas is needed for the calibration of the zero point. |
| Standard test gas | The standard test gas is required for calibration / adjustment of the sensor with a two-point calibration. The concentration value corresponds to about two-thirds of the full scale value of the probe up to 50% LEL. |
| Alarm test gas | The alarm test gas is required to verify the alarm levels. |
Select the measuring range of your probe:
0 - 50% LEL
| Measuring range of the probe | Gas | Usage | Bottle filling volume (Liter) | ||
|---|---|---|---|---|---|
| 34 l | 58 l | 110 l | |||
| 0 - 50% LEL | Synthetic air | Zero gas | Z0A0-L1373 | Z0A0-M1373 | Z0A0-N1373 |
| 0.75 vol% Butane in air | Standard test gas | Z0A0-L2252 | Z0A0-M2252 | Z0A0-N2252 | |
| 0.6 vol% Butane in air | Alarm test gas | Z0A0-L2248 | Z0A0-M2248 | Z0A0-N2248 | |
0 - 100% LEL
| Measuring range of the probe | Gas | Usage | Bottle filling volume (Liter) | ||
|---|---|---|---|---|---|
| 34 l | 58 l | 110 l | |||
| 0 - 100% LEL | Synthetic air | Zero gas | Z0A0-L1373 | Z0A0-M1373 | Z0A0-N1373 |
| 0.75 vol% Butane in air | Standard test gas | Z0A0-L2252 | Z0A0-M2252 | Z0A0-N2252 | |
| 0.6 Vol.-% Butane in air | Alarm test gas | Z0A0-L2248 | Z0A0-M2248 | Z0A0-N2248 | |
